Services List

 

 

  • Leasing Services

    You only need a system for a single project? Not sure how long you would need it? Not a problem. KONNIS can provide a liquid handling system with project programming for a short term lease. We offer 1,2,and 3 year lease terms for Tecan Liquid Handling systems

 



unique services




Volume Calibration.
on-site volume calibration of your Tecan Liquid Handler using a dye-water method or inetegrated balance

Request info

G4.2

(please send your instrument model)

 
Services Overview


On Site Automated Process Consultation.
with years of experience designin, installing, and implementing automated liquid handler systems let a KONNIS representative meet with you. We'll discuss your application specfic needs and work with you on the best soution for you.



Training.
Konnis can provide training on your Tecan Genesis, Genesis Freedom, TeMO, and Freedom EVO system. Konnis representative are fully trained on Tecan Software including(but not limited to): Gemini training, Evoware training, Evoware PLUS training, FACTS, Setup and Service, Logic, and Magellan.

 

 

Service.
KONNIS can provide preventative maintenance (PM) and service for your Tecan Liquid handling system. We have inventory of parts as well as manufacturing capabilities to keep your Tecan liquid handler up and running. Tecan Service is provided across the United States and Canada.


tecan service



Featured Tips and Tricks for TECAN Gemini Software


Invalid Operand when moving LiHA.
If you've ever received the error message "Invalid Operand" here are a few of the causes.

1. Having a plate carrier with 4 plate position in the Y-direction and telling the arm to pipette outside of its range. For example, telling tip 1 to move to well 96 of a plate on site 4 of the carrier would put the arm outside of its mechanical range. You would see the "Invalid Operand Error".

2. Having smaller syringes than 1.0mL and pipetting at speeds that exceed the maximum pipetting range for the dilutor motor will also produce this error. Typically it displays only when you run the protocol. The system goes along, may perform its aspiration and then prior to executing the dispense step you see "Invalid Operand".

First Dispense of Multidispense volume is too low
Many times this is related to the rack dimensions as well as the liquid level offset when liquid detecting. If you detect the liquid level and then start to track down as you aspirate your container dimensions need to be accurate or the tips will not stay submerged in the liquid. This will cause the tips to pull some air at the end of the aspiration draw. This air then gets transfered over to the first dispense (possibly the second depending on the volume.)



These items and much more are covered in the Tecan training offered by KONNIS.
 
 
 